Z-1829-2012Class II

Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with...

Philips North America LLC / initiated 2012-05-21 / Terminated / root cause: design

Philips Healthcare Inc. began a recall of Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with FlatDetector with swivel cable Product Codes: 708034... on . FDA classified it as Class II and gives the reason as "MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1829-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Multi Diagnost Eleva II with swivel cable Product Codes; 708032, 708036 Multi Diagnost Eleva with FlatDetector with swivel cable Product Codes: 708034, 708037 This Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/ universal system. General R/F, performed along with more specialized interventional applications. This includes the following general areas: Digestive system Skeletal system Urinary system Reproductive system Respiratory system Circulatory system Various . Arthrograms . Myelograms . Facet joint injections . Discography SialographyFluoroscopy, Radiography and Angiography
Product code
IZI System, X-Ray, Angiographic
Reason for recall
MultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement
Root cause tag
design
FDA root cause
Device Design
Action
Philips Healthcare Imaging issued Urgent Field Safety Notice dated May 21, 2012, to accounts informing an inspection will be performed of the unit. The Cable guiding plate will be replaced. If the cables show any damage the complete cable set wll also be replaced. No action is be taken by the customer. For further information customers should call 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Quantity in commerce
433 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Nationwide Foreign: Australia Austria Belgium Bharain Canada China Columbia Czech Republic Denmark Dutch Antilles Egypt Estonia Finland France Germany Hong Kong Ireland Israel Italy Latvia Lebanon Luxembourg Malaysia Mexico Netherlands New Zealand Norway Philippines Qatar Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syria Taiwan Thailand Turkey United Kingdom
Product codes and lots
Units with cable swivel delivered from 2007

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1829-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1829-2012Class IITerminatedVoluntary: Firm initiated