Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1897-2012Class II

Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc...

Ethicon, Inc / initiated 2012-04-03 / Terminated / root cause: packaging

Ethicon, Inc. began a recall of Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc., Somerville, NJ 08876 USA Manufactured by Ferrosan... on . FDA classified it as Class II and gives the reason as "The lyophilized thrombin component of the Surgiflo Hemostatic Matrix kit with Thrombin has the potential to be non-sterile due to an undetected air leak which occurred during the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1897-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon, Inc
Product
Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc., Somerville, NJ 08876 USA Manufactured by Ferrosan Medical Devices Ferrosan Medical Devices A/S Sydmarken 5, DK-2860 Soeberg, Denmark Mixed with sterile saline or thrombin solution is indicated for surgical procedures (except opthalmic) for hemostasis when control of capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical.
Product code
LMF Agent, Absorbable Hemostatic, Collagen Based
Reason for recall
The lyophilized thrombin component of the Surgiflo Hemostatic Matrix kit with Thrombin has the potential to be non-sterile due to an undetected air leak which occurred during the manufacturing process.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Ethicon Inc. sent a Urgent Voluntary Product Recall letter dated April 3, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you have any product subject to this voluntary recall, please remove it from your inventory immediately and return it per the following instructions. Return product to Stericycle using the enclosed pre-paid shipping label. Shipping fees wil be charged directly to Ethicon through the use of this shipping label. No reimbursements will be made for the shipping charge if this label is not used. Extra shipping labels may be obtained by calling 1-866-338-2611. Credit will be issued for impacted product that is returned. Further questions please call 1-866-338-2611, or contact your Ethicon Biosurgery Sales Representative. Thank you for your coorporation and immediate assistance. .
Quantity in commerce
27,396 kits
Distribution
USA (nationwide)
Product codes and lots
D100622   Product Code Lot Number Expiration Date  199102S 237115 8/31/2012 199102S 238568 9/30/2012 199102S 238569 11/30/2012 199102S 237493 6/30/2012 2993 235489 11/30/2012 2993 238599 3/31/2013 2993SAMP 238369 3/31/2013 SAMP199102S 237493 6/30/2012 SAMP199102S 283569 11/30/2012

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1897-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1897-2012Class IITerminatedVoluntary: Firm initiated