Z-0052-2013Class II

Sarns Centrifugal System Control module, 220/240V The Sarns Centrifugal System is indicated for use in...

Terumo Cardiovascular Systems Corporation / initiated 2012-09-24 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Sarns Centrifugal System Control module, 220/240V The Sarns Centrifugal System is indicated for use in cardiopulmonary bypass procedures only on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) received one report of loss of power to the control module and subsequently the drive motor for the Sarns centrifugal system during cardiopulmonary bypass". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0052-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns Centrifugal System Control module, 220/240V The Sarns Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.
Product code
KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) received one report of loss of power to the control module and subsequently the drive motor for the Sarns centrifugal system during cardiopulmonary bypass.
Root cause tag
design
FDA root cause
Component design/selection
Action
TERUMO sent an Urgent Medical Device Recall dated April 13, 2012, to all affected consignees. An Addendum was sent on September 25, 2012. All US Consignees will receive a Safety Advisory and an update to the Operator's Manual via Federal Express to inform them of the reported malfunction and potential hazards and to raise awareness of the benefits of using a retrograde valve and the importance of back-up system availability. TERUMO will include a customer response form and follow up with all affected users to obtain a response to the recall. For questions customers should call 1-800-521-2818. For questions regarding this recall call 734-741-6173.
Quantity in commerce
213 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Bangladesh, Seoul Korea, Chile, Brazil, Venesuela, Thailand, Mexico, India, Taiwan, Canada, Dominican Republic, Guatemala, Ecuador, Germany, Republic of Georgia, Hong Kong, Japan, United Arab Emirates (UAE), Japan, Belgium, Phillipines, Singapore, and South Africa.
Product codes and lots
Catalog number: 6380 and serial numbers: 5007, 5028-5030, 5032, 5075, 5076, 5135,6010,6013,6025,6030,6116, 7000-7197,7372, and 7373.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0052-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0051-2013Class IITerminatedVoluntary: Firm initiated
Z-0052-2013Class IITerminatedVoluntary: Firm initiated
Z-0053-2013Class IITerminatedVoluntary: Firm initiated