Intellispace Portal software. Intellispace Portal is a multimodality thin client applications server that...
Philips Medical Systems (Cleveland) Inc began a recall of Intellispace Portal software. Intellispace Portal is a multimodality thin client applications server that delivers full diagnostic viewing and clinical... on . FDA classified it as Class II and gives the reason as "Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration"...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1988-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Intellispace Portal software. Intellispace Portal is a multimodality thin client applications server that delivers full diagnostic viewing and clinical applications to the enterprise.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a feature that allows the user to launch images of multiple patient studies into an application, but does not display a warning message informing the user that diff
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated May 23, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact your local Philips representative or the Customer Care Solutions Center at 1-800-722-9377, option 5, for questions regarding this recall.
- Quantity in commerce
- 96
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, FL, GA, IA, IN, MI, NC, NJ, NY, OK, PA, SC, TN, TX, and WA and the countries of Australia, Austria, Canada, China, Czech Republic, Denmark, and France.
- Product codes and lots
- Model number 881001; Serial numbers: 81172, 79937, 85083, 81097, 85145, 79916, 81132, 85011, 82020, 85007, 85006, 85081, 81028, 80914, 82004, 85041, 81180, 82016, 30019, 80910, 81133, 81098, 81169, 79940, 80980, 82036, 82007, 85084, 85085, 82039, 81127, 81165, 30003, 80818, 30007, 85042, 81184, 82021, 82030, 82023, 82051, 82026, 82024, 82028, 82050, 82025, 82029, 82027, 30018, 79933, 82015, 85089, 30036, 81183, 82037, 80425, 81157, 85025, 82011, 81174, 80965, 80966, 81026, 35006, 82009, 81013, 82003 & 80891.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1988-2012 | Class II | Terminated | Voluntary: Firm initiated |