Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max...
Leica Microsystems, Inc. began a recall of Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage... on . FDA classified it as Class II and gives the reason as "Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2053-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Leica Microsystems, Inc.
- Product
- Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
- Product code
- FSO Microscope, Surgical, General & Plastic Surgery
- Reason for recall
- Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Leica Microsystems sent an Urgent Medical Device letter dated June 4, 2012 to the affected customers. The letter described the product, problem and actions to be taken. The customers were informed that the Leica surgical microscope M525 F50 required a software upgrade. Customers were instructed to NOT USE a remote video adapter or motorized video adapter until the required software upgrade has been installed by their Leica Microsystems representative. The customers were requested to complete the attached Acknowledgement Form, acknowledging receipt and understanding of the letter, listing the serial numbers of the units at their facilities and to fax it back to Leica Microsystems at (847) 236-3747.
- Quantity in commerce
- 39 Units - USA , 4 Units - Canada
- Distribution
- Worldwide Distribution - US (nationwide) in the states of: California, Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Michigan, Minnesota, New York, Oklahoma, Oregon, Pennsylvania, Texas, Utah, West Virginia, including Puerto Rico, and the countriy of Canada.
- Product codes and lots
- Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2053-2012 | Class II | Terminated | Voluntary: Firm initiated |