Z-1925-2012Class I

FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the...

Maquet Cardiovascular Us Sales, LLC / initiated 2012-02-20 / Terminated / root cause: design

Maquet Cardiovascular Us Sales, Llc began a recall of FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to... on . FDA classified it as Class I and gives the reason as "A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1925-2012
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
FLOW-I Anesthesia System The system is intended for use in administrating anesthesia while controlling the entire ventilation for patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
A technical alarm may be generated on the FLOW-i system when using the MAN/AUTO switch to change ventilation modes from manual to automatic or from automatic to manual if the switch is not fully engaged in the "on" or "off" position but remains in the "in-between position.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Maquet Getinge Group, sent an "URGENT DEVICE FIELD CORRECTION" letter dated February 20, 2012 to its customers. The letter describes the product, problem and action to be taken. The letter state that a Maquet Service Representative will upgrade the customers software in the FLOW-i Anesthesia System to the current revision and also provide the customers with the corresponding user's manual. Note: This upgrade includes other minor improvements. The customers were instructed to sign a document to verify satisfactory completion of the work once the software upgrade was completed. If you have questions or require additional information, please contact your local MAQUET representative or Technical Support at 888-627-8383. A press release was issued on 7/6/2012.
Quantity in commerce
284 units
Distribution
Worldwide distribution: USA (nationwide) state of GA only and countries of: Argentina, Australia, Austria, Belgium, China, Czech Republic, Ecuador, Finland, France, Germany, India, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Singapore, South Africa, Spain, Sweden incl Maquet Critical Care, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
Product codes and lots
Model Number :6677300 Serial Number : 1170

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1925-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1925-2012Class ITerminatedVoluntary: Firm initiated