Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue Plus(R) Pressure Injectable Central Venous Catheter...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2012-06-07 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue Plus(R) Pressure Injectable Central Venous Catheter - Catalog number CDC-42703-XP1A The multiple-lumen... on . FDA classified it as Class II and gives the reason as "Arrow is recalling the product due to the possibility of voids in the catheter hub which could lead to an interlumen crossover". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1823-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue Plus(R) Pressure Injectable Central Venous Catheter - Catalog number CDC-42703-XP1A The multiple-lumen pressure injectable catheter permits venous access to central circulation and injection of contrast media. The ARROWg+ard(R) antimicrobial surface is intended to provide protection against catheter-related infections.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Arrow is recalling the product due to the possibility of voids in the catheter hub which could lead to an interlumen crossover.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Arrow International sent a Urgent Medical Device Recall Notification letter dated June 11, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to take the following action: 1. If you have affected stock, immediately discontinue use and quarantine any products with the catalog and lot numbers listed. 2. To return product complete the enclosed Recall Acknowledgement Form and fax to 1-800-343-2935, Attn: Customer Service. This will allow us to document the amount of the product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization ( RGA) Number and will provide instructions for the return of product to Arrow International. 3. If you have no affected stock, please complete t he enclosed Recall Acknowledgment Form and fax it to 1-800-343-2935, Attn: Customer Service. This will allow us to document your receipt of this letter. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-800-343-2935.
Quantity in commerce
1,085
Distribution
USA (nationwide) AL, AZ, AR, CA, CO, FL, GA, IL, IN, KS, LA, MA, MI, MO, NV, NH, NC, OH, PA, TX, VA, WA, and WI.
Product codes and lots
Lot number RF2033361

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1823-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1823-2012Class IITerminatedVoluntary: Firm initiated