Intellispace Portal software
Philips Medical Systems (Cleveland) Inc began a recall of Intellispace Portal software Product Usage: Intellispace Portal is a multimodality thin-client applications server that delivers full diagnostic viewing and... on . FDA classified it as Class II and gives the reason as "In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2000-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Intellispace Portal software Product Usage: Intellispace Portal is a multimodality thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. Intellispace Portal is a medical software system that allows multiple users to remotely access the system from compatible computers on a network.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Healthcare sent an Urgent-Medical Device Correction letter dated May 23, 2012, via certified mail to each consignee. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the communication. The letter informed affected customers that a Philips Service Engineer will contact them for implementation of the software update on the affected systems free of charge. For any further information or support, customers can contact their local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompts).
- Quantity in commerce
- 44
- Distribution
- Worldwide Distribution - US (nationwide) including the states of: AR, LA, MI, NC, NY, OH, OK and PA and the countries of: Australia, Canada, Denmark, France, Germany, Israel, Netherlands, Spain and Switzerland.
- Product codes and lots
- Model number: 881001; serial numbers: 80818, 85088, 81178, 82006, 82038, 81063, 82018, 79935, 81150, 81013, 81173, 85068, 79932, 79943, 80911, 82019, 81027, 85049, 85043, 82008, 81119, 81089, 85108, 85026, 85146, 80994, 81021, 81182, 81017, 85067, 81161, 81164, 79936, 85103, 81147, 85110, 85082, 81179, 81035, 85046, 81031, 81129, 81004 & 85104.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2000-2012 | Class II | Terminated | Voluntary: Firm initiated |