Z-1933-2012Class II

PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery...

Zimmer GmbH / initiated 2012-06-05 / Terminated / root cause: design

Zimmer, Inc. began a recall of PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod... on . FDA classified it as Class II and gives the reason as "Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1933-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
PathFinder NXT Pivoting Percutaneous Rod Holder This device is an instrument used during orthopedic surgery. The intended use of the pivoting percutaneous rod holder is to attach to the rod to facilitate rod insertion, positioning, and orientation through the extender sleeves, under the fascia. The rod is secured to the holder by tightening a captive set screw against the rod end in the tip of the holder using the set screw driver. After positioning the rod and securing at least one pedicle screw closure top against the implanted rod, the holder is removed from the rod end by loosening the captive set screw with the set screw driver.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Complaints have been reported where, during surgery, the surgeon was unable to loosen or remove the inserted rod from the rod holder instrument. The complaint investigation showed the hex mating feature of the set screw was broken and material missing. There were no reports of fragments of the set screw being left in the patient or that there was any patient or user injury, but these events we
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Zimmer Spine, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 5, 2012 to all consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to review the notification and ensure they are aware of the content; to immediately remove and discontinue use of Pivoting Percutaneous Rod Holder (part number 3564-1) from Pathfinder NXT instrument sets in their possession and follow the PathFinder NXT Surgical Technique; to notify their Zimmer Spine customer service representative to obtain return authorization; and return the complete Surgeon Contact Certification Form within 10 working days from receipt of notification to [email protected] or Fax to 1-512-258-0995. If you have any questions, call Zimmer Spine Customer Service at 866-774-6368.
Quantity in commerce
77 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Australia, France and Germany.
Product codes and lots
Part number 3564-1; all lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1933-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1933-2012Class IITerminatedVoluntary: Firm initiated