Z-1984-2012Class II

Pudenz Flushing Valve with Integral Connectors, catalog number NL850-1330. Integra Neurosciences, Plainsboro...

Integra LifeSciences Corporation / initiated 2012-04-04 / Terminated / root cause: labeling and UDI

Integra LifeSciences Corp began a recall of Pudenz Flushing Valve with Integral Connectors, catalog number NL850-1330. Integra Neurosciences, Plainsboro, NJ. Device is used for the treatment of... on . FDA classified it as Class II and gives the reason as "The pouch label of catalog NL850-1330 indicates an incorrect model and closing pressure of the device. The incorrect label reads: "Model: 12 mm-Medium Pressure" instead of "Model: 16 mm-Low Pressure"". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1984-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Pudenz Flushing Valve with Integral Connectors, catalog number NL850-1330. Integra Neurosciences, Plainsboro, NJ. Device is used for the treatment of hydrocephalic patients when shunting cerebrospinal fluid from the ventricles of the brain.
Product code
JXG Shunt, Central Nervous System And Components
Reason for recall
The pouch label of catalog NL850-1330 indicates an incorrect model and closing pressure of the device. The incorrect label reads: "Model: 12 mm-Medium Pressure" instead of "Model: 16 mm-Low Pressure".
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The firm initiated a voluntary recall by notifying the two consignees initially via email and then with an "URGENT: MEDICAL DEVICE RECALL" dated April 5th or 9th, 2012. The letter informed them of the labeling error and provided instructions regarding affected product. The International Customer Service number was provided 609-936-5400.
Quantity in commerce
39 units
Distribution
International Only Distribution, including the countries of Thailand and Japan.
Product codes and lots
1115647, 1115939, 1120413

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1984-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1984-2012Class IITerminatedVoluntary: Firm initiated