Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX...
SALTER LABS began a recall of Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by... on . FDA classified it as Class II and gives the reason as "A complaint was received, and later verified, that product labeled as SO-1296 with male luer lock connector actually contained product with a female luer lock connector". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1898-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Salter Labs
- Product
- Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by Tri-Anium Assembled in Mexico. Usage: Delivery of Anesthesia.
- Product code
- CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- Reason for recall
- A complaint was received, and later verified, that product labeled as SO-1296 with male luer lock connector actually contained product with a female luer lock connector.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- The firm, Salter Labs, called their customers to discuss the recall and followed up with a "PRODUCT RECALL" letter dated May 29, 2012. The letter was distributed to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory and return any of the lots in their possession to Salter Labs; contact Salter Labs Customer Service at 1-800-235-4203 to arrange for return of the product, and complete and return the attached form/record via fax to: 661-854-6816 or Toll Free 1-800-628-4690. Should you have any questions, please contact Salter Labs at 1-800-235-4203.
- Quantity in commerce
- 242 cases
- Distribution
- Nationwide distribution: USA including states of: FL, IN, KS, PA and TX.
- Product codes and lots
- Salter Part Number SO-1296, Tri-Anium Part numaber: 301-P3600; Lot number: 032612.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1898-2012 | Class II | Terminated | Voluntary: Firm initiated |