Gram Crystal Violet 250 mL, catalog #212525, in cartons of 4 bottles. Gram Stain Kits and Reagents are used...
Becton Dickinson & Co. began a recall of Gram Crystal Violet 250 mL, catalog #212525, in cartons of 4 bottles. Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens... on . FDA classified it as Class II and gives the reason as "Potential for increase level of artifacts to appear as large, spherical and easy to distinguish or small & more difficult to distinguish from gram positive cocci". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2032-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Gram Crystal Violet 250 mL, catalog #212525, in cartons of 4 bottles. Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.
- Product code
- JTS Stains, Microbiologic
- Reason for recall
- Potential for increase level of artifacts to appear as large, spherical and easy to distinguish or small & more difficult to distinguish from gram positive cocci.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard the affected product. An Acknowledgement Form was attached for customers to complete and return via fax to 410-316-4258. Contact the BD Customer Service Department for questions regarding this recall.
- Quantity in commerce
- 878 cartons/ 3512 bottles
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lots 1325324 Exp May 31,2013 and 2011746, Exp July 31,2013
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2031-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2032-2012 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.