Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2198-2012Class II

Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model TS057HF7. For use in patients who require...

Edwards Lifesciences LLC / initiated 2012-05-11 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model TS057HF7. For use in patients who require hemodynamic monitoring on . FDA classified it as Class II and gives the reason as "The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2198-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model TS057HF7. For use in patients who require hemodynamic monitoring.
Product code
DYG Catheter, Flow Directed
Reason for recall
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Edwards Lifesciences sent a "PRODUCT RECALL" letter dated May 8, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Confirmation Form was attached for customers to complete and return via fax to 1-800-422-9329. Contact Customer Service at 1-800-424-3278 for questions regarding this notice.
Quantity in commerce
32,145 units for all products in Recall Event
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Japan, EMEA (Europe, Middle East and Africa), Asia Pacific, (Hong Kong, Indonesia, Malaysia, Australia, and Taiwan, Thailand), Latin America, (Chile, Colombia, Panama, Mexico, and Honduras).
Product codes and lots
59267057

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2198-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2188-2012Class IITerminatedVoluntary: Firm initiated
Z-2189-2012Class IITerminatedVoluntary: Firm initiated
Z-2190-2012Class IITerminatedVoluntary: Firm initiated
Z-2191-2012Class IITerminatedVoluntary: Firm initiated
Z-2192-2012Class IITerminatedVoluntary: Firm initiated
Z-2193-2012Class IITerminatedVoluntary: Firm initiated
Z-2194-2012Class IITerminatedVoluntary: Firm initiated
Z-2195-2012Class IITerminatedVoluntary: Firm initiated
Z-2196-2012Class IITerminatedVoluntary: Firm initiated
Z-2197-2012Class IITerminatedVoluntary: Firm initiated
Z-2198-2012Class IITerminatedVoluntary: Firm initiated