Hudson RCI Concha Therm Neptune Humidifier, Rx Only
Teleflex Medical began a recall of Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing... on . FDA classified it as Class II and gives the reason as "Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2036-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Hudson RCI Concha Therm Neptune Humidifier, Rx Only, Product Usage: Usage: Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.
- Product code
- BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
- Reason for recall
- Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Teleflex sent Urgent Medical Device Notification letters dated June 12, 2012 to all affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to complete the enclosed Acknowledgement Form and return per the instructions provided. Once the form is received, Teleflex Medial will schedule for a technician to come to their facility and exchange the affected product. For questions call your local sales representative or Customer Service at 1-866-246-6990.
- Quantity in commerce
- 14,794 each
- Distribution
- Worldwide Distribution - U.S. (nationwide) and the countries of Argentina, Australia, Brazil, Canada, Chile, China, Europe, India, Japan, Mexico, South Korea, Taiwan and UK.
- Product codes and lots
- Catalog #425-00, 425-10 and 425-30.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2036-2012 | Class II | Terminated | Voluntary: Firm initiated |