Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2202-2012Class II

Carestream DRX-1 System Battery --- Made in U.S.A. for Carestream Health, Inc., Carestream Health, Inc., 150...

Eastman Kodak Company / initiated 2012-06-12 / Terminated / root cause: manufacturing process

Carestream Health, Inc. began a recall of Carestream DRX-1 System Battery --- Made in U.S.A. for Carestream Health, Inc., Carestream Health, Inc., 150 Verona St, Rochester, NY 14608. --- The battery... on . FDA classified it as Class II and gives the reason as "Carestream DRX-1 System batteries manufactured before June 2012 due to two occurrences of lithium-ion battery overheating, which may cause burns". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2202-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
Carestream DRX-1 System Battery --- Made in U.S.A. for Carestream Health, Inc., Carestream Health, Inc., 150 Verona St, Rochester, NY 14608. --- The battery may be sold with the DRX-1 imaging device or as a stand-alone accessory.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Carestream DRX-1 System batteries manufactured before June 2012 due to two occurrences of lithium-ion battery overheating, which may cause burns.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Firm sent consignees "Urgent: Customer Notification and Field Corrective Action" letters on June 8, 2012. The letters described the issue and provided recommended actions. Service dealers will contact all affected consignees to evaluate the battery. U.S. customers with questions can contact Carestream Customer Care at 800-328-2910. Customers outside the United States should contact their local Carestream service number.
Quantity in commerce
9971 units
Distribution
Worldwide Distribution, including Nationwide (USA).
Product codes and lots
All DRX-1 batteries manufactured before June 2012.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2202-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2202-2012Class IITerminatedVoluntary: Firm initiated