Z-0128-2013Class II

Perfusion System 8000, base 4 pump, 220V/240V safety monitor included The Sarns Modular Perfusion System 8000...

Terumo Cardiovascular Systems Corporation / initiated 2012-09-25 / Terminated / root cause: labeling and UDI

Terumo Cardiovascular Systems Corporation began a recall of Perfusion System 8000, base 4 pump, 220V/240V safety monitor included The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the... on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Modular Perfusion System 8000. Very few of...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0128-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Perfusion System 8000, base 4 pump, 220V/240V safety monitor included The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Modular Perfusion System 8000. Very few of the reports occurred during cardiopulmonary bypass; the majority of the malfunctions occurred during set-up or were reported during routine preventative maintenance. The malfunction
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
TERUMO sent an URGENT MEDICAL DEVICE letter dated September 25, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed review the Medical Device Recall notice, assure that all users are aware of the notice and place the addendum in the Operator's Manual before page 1.13. Customers were also instructed to confirm receipt of the Urgent Medical Device letter by faxing, the attached Customer Response Form to 1-800-292-6551. For questions customers were instructed to call 1-800-521-2818.
Quantity in commerce
466 units
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY and the countries of ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, Bolivia, BRAZIL, CALLO PERU, CANADA, CAROLINA PR, CHILE, China, COLOMBIA, Costa Rica, DEKALB, DENMARK, Dominican Republic, Ecuador, Egypt, FRANCE, Germany, GUAYQUIL ECUADOR, Hong Kong, India, Indonesia, ITALY, Japan, , JARAMA MADRID, Lebanon, LEICESTERSHIRE, MADRID SPAIN, MADRID SPAIN, Malaysia, MEDLEY, Mexico, D.F., MIAMI, MUMBAI 26 INDIA, MUMBAI, New Zealand, Nicaragua, Pakistan, Panama, Philippines, REPUBLICA, SANTIAGO, Saudi Arabia, SELANGOR, MALAYSA, SEOUL, KOREA, Singapore, SOUTH AFRICA, South Korea, Sri Lanka, SWITZERLAND, TAIPEI, TAIWAN, Thailand, TUJUNGA, UNITED ARAB EMIRATES (UAE), URUGUAY, Venezuela, Vietnam, VIGNATE ITALY, WOODDALE, and Yemen.
Product codes and lots
Catalog number: 16405 and serial numbers: 1001-1466

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0128-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0126-2013Class IITerminatedVoluntary: Firm initiated
Z-0127-2013Class IITerminatedVoluntary: Firm initiated
Z-0128-2013Class IITerminatedVoluntary: Firm initiated
Z-0129-2013Class IITerminatedVoluntary: Firm initiated
Z-0130-2013Class IITerminatedVoluntary: Firm initiated
Z-0131-2013Class IITerminatedVoluntary: Firm initiated