Z-2348-2012Class II

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use...

Terumo Cardiovascular Systems Corporation / initiated 2009-08-12 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation... on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2348-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The safety monitor is indicated to warn when the blood level in the reservoir reaches the operator-set limit, and to stop the arterial and cardioplegia pumps.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav
Root cause tag
manufacturing process
FDA root cause
Mixed-up of materials/components
Action
TERUMO Cardiovascular Systems communicated with their affiliate, Terumo Dubai, in 2009 at the time the complaints were reported. No further communication was warranted. For questions regarding this recall call 734-741-6173.
Quantity in commerce
8
Distribution
Worldwide distribution: International only-country of: Terumo Dubai, UAE
Product codes and lots
catalog number 16404 and serial numbers: 4357, 4381, 4386, 4387, 4389, 4391, and 4392.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2348-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2348-2012Class IITerminatedVoluntary: Firm initiated