Z-2449-2012Class II

Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for...

Terumo Cardiovascular Systems Corporation / initiated 2012-09-20 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and... on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2449-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1 Flow Module indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat
Root cause tag
manufacturing process
FDA root cause
Process control
Action
TERUMO sent an Urgent Medical Device Recall letter dated September 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the Medical Device Recall notice; assure that all users are aware of the notice; and confirm receipt of the communication by faxing the attached Customerr Resonse Form to 1-734-741-6149. Terumo CVS will send a representative to each affected US account to replace the affected devices and return the affected devices to Terumo for destruction. Non-US Terumo Affiliates will be instructed to remove the devices from the field and provide a record of destruction. For question customers should call 1-800-521-2818. For questions regarding this recall call 734-741-6173.
Quantity in commerce
72 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of AUSTRALIA, Singapore, Jordan, Japan, Vietnam, Germany, Russia, Saudi Arabia, BELGIUM, Mexico, and CANADA.
Product codes and lots
Catalog number: 802018 and serial numbers: 00122,00132, 00137, 00321-00337, 00339, 00340, 00342-00389,00391-00439, 00441, 00442, 00444-00455, 00457, 00459, 00460, 00466, 00467, 00475, and 00488.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2449-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2449-2012Class IITerminatedVoluntary: Firm initiated