Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1980-2012Class II

Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone...

Zimmer GmbH / initiated 2012-06-05 / Terminated / root cause: design

Zimmer, Inc. began a recall of Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The... on . FDA classified it as Class II and gives the reason as "Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1980-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
Zimmer sent an Urgent Medical Device Correction and Removal notification dated June 5, 2012 to distributors and all affected accounts of Zimmer Spine. The notice identified the affected products, problem and actions to be taken. Distributors were asked to hand deliver Recall Notices to Surgeons and return the completed Surgeon Contact Certification Form within 10 working days to [email protected] or FAX to 1-512-258-0995. Notices to Surgeons and User facilities included revised directions/precautions for the Instructions for Use and attached Surgical Techique. Devices will be removed once newly design parts become available. For any related questions or assistance about this please contact Zimmer Spine Customer Service at 866-774-6368.
Quantity in commerce
249 ( angle handle and straight) in total
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
Product codes and lots
Part numbers 3573-1 (angle handle) all lots including 64PY, 67YA, 69FF, 71GB, 72JL, 73GY

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1980-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1980-2012Class IITerminatedVoluntary: Firm initiated
Z-1981-2012Class IITerminatedVoluntary: Firm initiated