Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone...
Zimmer, Inc. began a recall of Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The... on . FDA classified it as Class II and gives the reason as "Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1980-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Zimmer sent an Urgent Medical Device Correction and Removal notification dated June 5, 2012 to distributors and all affected accounts of Zimmer Spine. The notice identified the affected products, problem and actions to be taken. Distributors were asked to hand deliver Recall Notices to Surgeons and return the completed Surgeon Contact Certification Form within 10 working days to [email protected] or FAX to 1-512-258-0995. Notices to Surgeons and User facilities included revised directions/precautions for the Instructions for Use and attached Surgical Techique. Devices will be removed once newly design parts become available. For any related questions or assistance about this please contact Zimmer Spine Customer Service at 866-774-6368.
- Quantity in commerce
- 249 ( angle handle and straight) in total
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
- Product codes and lots
- Part numbers 3573-1 (angle handle) all lots including 64PY, 67YA, 69FF, 71GB, 72JL, 73GY
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1980-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1981-2012 | Class II | Terminated | Voluntary: Firm initiated |