Z-2077-2012Class II

.***ANSPACH***Sterile. REF SHD-4820DS***QTY: 1, 4.8 mm x 2mm Twist Drill with Depth Stop. Use with SHORT-HD...

The Anspach Effort, Inc. / initiated 2011-09-19 / Terminated / root cause: design

The Anspach Effort, Inc. began a recall of .***ANSPACH***Sterile. REF SHD-4820DS***QTY: 1, 4.8 mm x 2mm Twist Drill with Depth Stop. Use with SHORT-HD Attachment ***RX Only Product Usage: Cutting and... on . FDA classified it as Class II and gives the reason as "Anspach Effort, Inc. Palm Beach Gardens, FL is recalling their Single Use Sterile Bone Cutting Burrs used with XMax, microMax, microMax Plus, eMax, eMax2 and eMax2 Plus High Speed Drill Systems. It is possible for the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2077-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Anspach Effort, Inc.
Product
.***ANSPACH***Sterile. REF SHD-4820DS***QTY: 1, 4.8 mm x 2mm Twist Drill with Depth Stop. Use with SHORT-HD Attachment ***RX Only Product Usage: Cutting and shaping bone including bones of the spine and cranium.
Product code
HTW Bit, Drill
Reason for recall
Anspach Effort, Inc. Palm Beach Gardens, FL is recalling their Single Use Sterile Bone Cutting Burrs used with XMax, microMax, microMax Plus, eMax, eMax2 and eMax2 Plus High Speed Drill Systems. It is possible for the cutter to come apart.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Anspach, sent an "URGENT: Medical Device Product Removal" letter dated December 15, 2011, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Screen their inventory immediately for any of the batches listed in the letter. 2) If they have any of the products listed, contact Anspach Customer Support at (800) 327-6887 to arrange for their immediate return and replacement. 3) Complete and return the attached Customer Reply form indicating if the item(s) are being returned and confiming their receipt of this letter, and fax to 1-800-327-6661. If they have distributed any of the products to other services or facilities, please forward this information as appropriate. Should you have any queries, please do not hesitate to contact Anspach Customer Support at (800) 327-6887.
Quantity in commerce
5 each
Distribution
US Nationwide Distribution including the states of: Arizona. and California.
Product codes and lots
Lot numbers E203050779 and E383055709.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2077-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2077-2012Class IITerminatedVoluntary: Firm initiated