Z-1896-2012Class II
Philips HeartStart MRx monitor/defibrillators Models: M3535A, M3536A, M3536M4, M3536M5, and M3536M6
Philips North America LLC / initiated 2012-06-20 / Terminated / root cause: manufacturing process
Philips Healthcare Inc. began a recall of Philips HeartStart MRx monitor/defibrillators Models: M3535A, M3536A, M3536M4, M3536M5, and M3536M6 Product Usage: The Heartstart MRx is for use for the... on . FDA classified it as Class II and gives the reason as "HeartStart MRx Unexpected Pads/Paddles ECG Failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1896-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips HeartStart MRx monitor/defibrillators Models: M3535A, M3536A, M3536M4, M3536M5, and M3536M6 Product Usage: The Heartstart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- HeartStart MRx Unexpected Pads/Paddles ECG Failure
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Philips issued an Urgent Voluntary Medical Device Correction/Field Safety Notice letter on June 20, 2012. The letter informs customers of the issue, identifies details (including models and software versions applicable) of units affected, gives instructions on actions to be taken until the correction is available, and identifies what action Philips plans to eliminate or remedy the issue. The letter states that a Philips Healthcare representative will contact customers to arrange for the replacement of the Power PCA in all the affected devices. Philips will complete the installation free of charge. For questions or support concerning this issue contact your local Philips representative at 1-800-722-9377.
- Quantity in commerce
- 2914 units
- Distribution
- Worldwide Distribution - US (nationwide) and the countries of: AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, COLOMBIA, EL SALVADOR, ETHIOPIA, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOREA, REPUBLIC OF, LEBANON, LIBYAN ARAB JAMAHIRIYA, MEXICO, MOROCCO, NETHERLANDS, NETHERLANDS ANTILLES, NICARAGUA, NORWAY, PERU, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, and VIET NAM.
- Product codes and lots
- Manufactured between the dates of July 12, 2011 through September 2, 2011 and January 19, 2012 through May 22, 2012. Serial numbers within the ranges of: US00550047 to US00551654 and US00556357 to US00559493 respectively. There are also some additional MRx units that were repaired and may contain compromised components. These MRx unit serial numbers are identified below: US00210180, US00210183, US00210738, US00211000, US00213281 US00214706, US00316827, US00318163, US00319241, US00328491 US00329811, US00333101, US00536832, US00538140, US00543161 US00543825, US00544081, US00544214, US00546107, US00548046
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1896-2012 | Class II | Terminated | Voluntary: Firm initiated |