CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical...
CareFusion 203, Inc. began a recall of CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is... on . FDA classified it as Class I and gives the reason as "CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault. The unit will activate...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2006-2012
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Pulmonetic Systems, Inc.
- Product
- CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. It is not intended for homecare use.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault. The unit will activate either a Low PEEP or Patient Circuit Fault Alarm both audibly and Visually to alert the healthcare professional. Ventilation delivery to the patient may be compromised.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- CareFusion sent an Urgent Product Recall letter with a return receipt dated June 19, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Required Action for Users: If the ventilator resumes normal ventilation (intermittent leak) the audible alarm stops. To clear the alarm indicator on the ventilator display, enter Alarm Messages tab, push alarm Reset to clear display. If the ventilator does not resume normal ventilation (continuous leak) provde an alternative method of ventilation to the patient. CareFusion will follow-up with a phone call to coordinate the correction of the devices. For questions customers should call 1-800-554-8933.
- Quantity in commerce
- 266
- Distribution
- Nationwide Distribution including AZ, CA, CO, FL, IL, KY, ME, MI, MO, MN, MT, ND, NJ, NM, NV, NY, OK, PA, SC, SD, TX , UT, WI, and WV.
- Product codes and lots
- serial numbers: 26562, 43634, 43642, 43643, 44536, 44537, 44540, 44541, 44542, 44543, 44544, 44545, 44546, 44552, 44553, 44554, 44555, 44557, 44558, 44559, 44560, 44561, 44562, 44563, 44564, 44565, 45318, 45320, 45321, 45322, 45323, 45324, 45325, 45326, 45327, 45328, 45329, 45330, 45333, 45334, 45335, 45336, 45337, 45338, 45339, 45340, 45341, 45342, 45343, 45344, 45345, 45346, 45735, 45736, 45737, 45738, 45739, 45740, 45741, 45742, 45743, 45744, 45745, 45746, 45748, 45750, 45751, 45752, 45753, 45754, 45902, 45903, 45904, 45905, 45906, 45907, 45908, 45909, 45910, 45911, 46038, 46039, 46040, 46041, 46042, 46043, 46044, 46045, 46046, 46047, 46048, 46049, 46050, 46051, 46052, 46632, 46634, 46635, 46636, 46637, 46638, 46639, 46640, 46641, 46642, 46643, 46644, 46645, 47069, 47070, 47071, 47073, 47074, 47075, 47076, 47077, 47078, 47079, 47080, 47081, 47082, 47083, 47210, 47211, 47212, 47213, 47214, 47215, 47216, 47217, 47219, 47221, 47222, 47223, 47225, 47226, 47227, 47229, 47230, 47231, 47232, 47233, 47234, 47236, 47237, 47238, 47239, 47538, 47539, 47540, 47541, 47542, 47543, 47544, 47545, 47546, 47547, 47548, 47549, 47550, 47551, 47552, 47553, 47554, 47555, 47556, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47565, 47566, 47567, 48759, 48760, 48761, 48762, 48763, 48764, 48765, 48766, 48767, 48768, 48769, 48770, 48771, 48772, 48773, 48774, 48775, 48776, 48777, 48778, 48779, 48780, 48781, 48782, 48783, 48784, 48785, 48786, 48787, 48788, 48868, 48869, 48870, 48871, 48872, 48873, 48874, 48875, 48876, 48878, 48881, 48883, 48884, 48885, 48886, 48887, 48889, 48890, 48891, 48892, 48893, 48894, 48895, 48897, 50390, 50391, 50392, 50393, 50394, 50395, 50397, 50409, 50411, 50412, 50416, 50417, 50418, 50419, 50648, 50650, 50651, 50652, 50654, 50656, 50657, 50658, 50659, 50660, 50661, 50662, 50663, 50664, 50665, 50666, 50667, 50668, 50672, 51569, 51574
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2006-2012 | Class I | Terminated | Voluntary: Firm initiated |