Z-2037-2012Class II

Hitachi Scenaria CT System Software

FUJIFILM Healthcare Americas Corporation / initiated 2012-05-04 / Terminated / root cause: software

Hitachi Medical Systems America Inc began a recall of Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images... on . FDA classified it as Class II and gives the reason as "Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2037-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.
Root cause tag
software
FDA root cause
Software design
Action
Hitachi sent a "Device Correction Notice" dated May 4, 2012 to their customers. The notice identified the affected product, problem and actions to be taken. Customers were instructed to "not use the Reconstruction Status" function to suspend active reconstructions until the software problem can be corrected. They also informed their customers that a representative will visit their facility to install the software update to permanently resolve the issue at no cost to the facility. For further information regarding this notice call 800-800-3106 x3720.
Quantity in commerce
Six (6) units
Distribution
US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.
Product codes and lots
Serial #'s S5001 - S5011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2037-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2037-2012Class IITerminatedVoluntary: Firm initiated