Respironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or...
Respironics, Inc. began a recall of Respironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of... on . FDA classified it as Class I and gives the reason as "The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2024-2012
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Respironics, Inc.
- Product
- Respironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy 100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The devices are intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. They are not intended to be used as a transport ventilator.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Respironics, Inc. began contacting its affected US customers by phone beginning on Friday, April 27, 2012, and requested that they quarantine any affected devices within their possession and that they retrieve any devices with patients and quarantine those as well. Replacement devices were to be provided to all affected customers. For questions customers should call 724-387-7651. For questions regarding this recall call 877-387-3311..
- Quantity in commerce
- 173
- Distribution
- Worldwide Distribution - USA including AZ, CA, IL, IN, LA, MI, MN, MO, NJ, NY, NC, PA, SC, TN, TX, and WI and the countries of Canada, Denmark, Estonia, Finland, France, Great Britian, India, Italy, Japan, Korea, Saudi Arabia, and Spain.
- Product codes and lots
- Model/Catalog Numbers: 1032800, 1040004, 1040007, 1054096, 1054260, CA1032800, R1040004, R1054655, and 1045151.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2024-2012 | Class I | Terminated | Voluntary: Firm initiated |