Z-2173-2012Class II

Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and...

Siemens Healthcare Diagnostics Inc. / initiated 2012-06-07 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing module for use with In-vitro diagnostic... on . FDA classified it as Class II and gives the reason as "The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2173-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments
Product code
NSX
Reason for recall
The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the LIS an incorrect result following the syngo LDM's application of the configured result number format. The problem is encountered when the result number format configured on the Sys
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
Siemens Healtcare Diagnostics sent a Urgent Field Safety Notice dated June 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised to discuss the contents of this letter with your Laboratory Director regarding the need to review any previous test results. If you have any technical questions or concerns, please contact the Seimens Technical Solutions Center at 1-800-441-9250, for further assistance.
Quantity in commerce
15
Distribution
USA (nationwide) including the states of AZ, CA, MS, MO, NJ, NC, OH, OR, PA, and TX.
Product codes and lots
Version VA11B and all previously released versions.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2173-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2173-2012Class IITerminatedVoluntary: Firm initiated