Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and...
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing module for use with In-vitro diagnostic... on . FDA classified it as Class II and gives the reason as "The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2173-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens syngo(R) Lab Data Manager. The syngo(R) Lab Data Manager is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments
- Product code
- NSX
- Reason for recall
- The recalling firm has discovered that when using a Sysmex(R) CA-7000 or CA-1500 coagulation analyzer with the syngo Lab Data Manager (LDM), there is a potential for the syngo LDM system, to display and transmit to the LIS an incorrect result following the syngo LDM's application of the configured result number format. The problem is encountered when the result number format configured on the Sys
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- Siemens Healtcare Diagnostics sent a Urgent Field Safety Notice dated June 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised to discuss the contents of this letter with your Laboratory Director regarding the need to review any previous test results. If you have any technical questions or concerns, please contact the Seimens Technical Solutions Center at 1-800-441-9250, for further assistance.
- Quantity in commerce
- 15
- Distribution
- USA (nationwide) including the states of AZ, CA, MS, MO, NJ, NC, OH, OR, PA, and TX.
- Product codes and lots
- Version VA11B and all previously released versions.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2173-2012 | Class II | Terminated | Voluntary: Firm initiated |