Stryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument...
Stryker Spine began a recall of Stryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument, sterilized prior to use. It allows the surgeon to... on . FDA classified it as Class II and gives the reason as "Stryker Spine has received six reports regarding the breakage of one or more prongs at the extremity of the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2176-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument, sterilized prior to use. It allows the surgeon to distract the disc space in the lumbar region of the spine using an anterior or anterolateral approach while also allowing them to insert the implant into final position.
- Product code
- MDM Instrument, Manual, Surgical, General Use
- Reason for recall
- Stryker Spine has received six reports regarding the breakage of one or more prongs at the extremity of the instrument.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Stryker sent an Urgent Product Recall letters/return response forms dated April 5, 2012, via FedEX to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and hospital location to identify the product. Reconcile any SLIDE Implant Blockers with catalog and lot numbers referenced above. Retrieve and return the affected product using the enclosed pre-paid mailing label to Stryker Spine 59 Route 17 South, Allendale, New Jersey, 07401. For questions customers were instructed to call 201-760-8298.
- Quantity in commerce
- 109 Lots
- Distribution
- Nationwide Distribution including NC, AR, TN, MA, IL, GA, NY, LA, MN, MO, OH, KS, NM, TX, KS,and IN.
- Product codes and lots
- Catalog Number: 48361223, 48361224, 48361225, 48361226, 48361227, 48361228 Lot Code: 084505, 084506, 084507, 084508, 084509, 084510
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2176-2012 | Class II | Terminated | Voluntary: Firm initiated |