Z-2002-2012Class II
Focal SIM Planning of radiation therapy
Elekta Limited / initiated 2012-06-21 / Terminated / root cause: software
Elekta, Inc. began a recall of Focal SIM Planning of radiation therapy on . FDA classified it as Class II and gives the reason as "Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2002-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Focal SIM Planning of radiation therapy
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- IMPAC Medical Systems, Inc. sent an Important Safety Notice LINFFM0001/1.0 on June 21, 2012, to consignees on concerning a patch to the software. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to not use the DICOM coordinates option for Reports when printing the Setup Reference Diaglog. Customers were instructed go to www.elekta.com, select the SupportPlus Login at the tope of the page, enter their portal, select Downloads/Updates to download the latest patches. For questions regarding this recall call 770-670-2548.
- Quantity in commerce
- 38 units
- Distribution
- Nationwide Distribution including CA, CT, FL, IL, IN, KY, MI, MN, MO, NJ, NY, NC, OH, OK, OR, PA, WA, and WI.
- Product codes and lots
- Version 4.64.00 - 4.700, inclusive
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2002-2012 | Class II | Terminated | Voluntary: Firm initiated |