Z-2007-2012Class II

Monaco

Elekta Limited / initiated 2012-06-21 / Terminated / root cause: software

Elekta, Inc. began a recall of Monaco Product Usage: Planning of radiation therapy on . FDA classified it as Class II and gives the reason as "Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2007-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco Product Usage: Planning of radiation therapy
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.
Root cause tag
software
FDA root cause
Software design
Action
Elekta sent an Important Safety Notice to affected customers. The notice identified the affected product, problem, clinical impact and actions to be taken. The notice inform customers of how the issue will be resolved and instructions to follow when it is available.
Quantity in commerce
38
Distribution
US Nationwide Distribution - including the states of: CA, CT, FL, IL, IN, KY, MI, MN, MO, NJ, NY, NC, OH, OK, OR, PA, WA, and WI
Product codes and lots
Version 3.10.00 - 3.20.00, inclusive

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2007-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2007-2012Class IITerminatedVoluntary: Firm initiated