UroView 2800; MDL number A349855
GE OEC Medical Systems, Inc began a recall of UroView 2800; MDL number A349855. Product Usage: Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and... on . FDA classified it as Class II and gives the reason as "Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2127-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- UroView 2800; MDL number A349855. Product Usage: Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical application may include but are not limited to urologic and endoscopic procedures. The system may be ued for other imaging applications at the physician's discretion.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC sent an Urgent Safety Advisory Notices dated October 11, 2006 and February 21, 2007 via certified mail delivery letter to affected customers. The notices identified the affected products, problem and actions to be taken. Customers were informed that a GEHC OEC field service engineer would be contacting their facility to arrange to have a replacement hard disk drive installed on the affected system(s). Customers were instructed to ensure that images on the hard disk drive were stored via an alternate permanent media as the hard disk drive approaches the 170 image capacity until a GEHC OEC field service engineer had replaced the drive. For questions regardiug this issues contact the service team at 800-874-7378.
- Quantity in commerce
- Part of a total of 82 units.
- Distribution
- Worldwide Distribution - US (nationwide) and the countries of: Canada, France, Japan, and China.
- Product codes and lots
- 62-0212 , 69-0054 , 82-0023-RC , 82-0034 , 82-0072 , 82-0274 , 82-0587, 8S-0668 , 82-1710 , 82-1803 , 82-2147 , 82-2620 , 82-2754 , 82-2815 , 82-3014 , 82-3234 , 82-7010-MH , 86-0254 , 86-0354 , 86-1078 , 89-0505 , 89-0509 , 89-0595 , 89-0605 , 89-0987 , 89-1048 , 89-1776 , 89-1837 , 89-2132 , 89-2303 , 89-2335 , 89-2348 , 89-2350 , 89-2610 , 89-2645 , 8S-0066 , 8S-0079 , 8S-0341 , 8S-0562 , 8S-0615-C , 8S-0847 , 8S-0849-C , 8S-1312 , 8S-1971 , 8S-2015-N , 8S-2116 , 8S-2117 , 8S-2199 , 8S-2204 , 8S-2320 , 8S-2411 , P4-0097-R , P4-0101-R , P4-0325-R , P4-0346-L , P4-0477-L , P6-0432-R , W2-0012 , W2-0580-R , W2-0591-R , 82-0030
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2125-2012 | Class II | Terminated | FDA Mandated | ||
| Z-2126-2012 | Class II | Terminated | FDA Mandated | ||
| Z-2127-2012 | Class II | Terminated | FDA Mandated |