Z-2074-2012Class II

BridgePoint Medical, Stingray Catheter, REF M - 1000, Sterilized with Ethylene Oxide Gas, Prescription Only...

Boston Scientific Corporation / initiated 2012-06-26 / Terminated / root cause: manufacturing process

Bridgepoint Medical began a recall of BridgePoint Medical, Stingray Catheter, REF M - 1000, Sterilized with Ethylene Oxide Gas, Prescription Only, Manufacturer BridgePoint Medical Inc. 13355 10th... on . FDA classified it as Class II and gives the reason as "BridgePoint Medical Inc. is voluntarily recalling Stingray Catheter, Model M1000 lots BP20121440078 and BP20121560089. BridgePoint Medical, Inc. recently determined the potential of compromised sterility assurance of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2074-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
BridgePoint Medical, Stingray Catheter, REF M - 1000, Sterilized with Ethylene Oxide Gas, Prescription Only, Manufacturer BridgePoint Medical Inc. 13355 10th Ave N, Suite 110, Plymouth, MN 55441. Stingray" Catheters are intended to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions) prior to PTCA or stent intervention. When used as part of the BridgePoint Medical System, the Stingray Catheter is intended to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions [CTOs]) prior to PTCA or stent intervention. .
Product code
DQY Catheter, Percutaneous
Reason for recall
BridgePoint Medical Inc. is voluntarily recalling Stingray Catheter, Model M1000 lots BP20121440078 and BP20121560089. BridgePoint Medical, Inc. recently determined the potential of compromised sterility assurance of the product due to a faulty pouch seal. In no case was there any reported adverse clinical event related to the problem.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, BridgePoint Medical, sent an "Urgent Field Safety Notice" letter dated June 26, 2012 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately examine their inventory, quarantine the affected lots and complete and return the enclosed "Verification Form" via fax to 763-225-8718 ,attn: BridgePoint Medical, Inc., Upon receipt of form, BridgePoint will supply the customers with the appropriate materials to return the products. If you have any questions regarding this notice, please call BridgePoint Medical, Vice President of Quality Assurance, at 763-225-8511 or e-mail at [email protected].
Quantity in commerce
95 (57 USA, 38 OUS)
Distribution
Worldwide distribution: USA (nationwide) including: AL, AZ, CA, CO, GA, IL, KY, MD, MN, MO, NC, NY, OR, PA, WA, and WI; and countries of: Australia and Sweden.
Product codes and lots
Lots affected in US: BP20121440078, BP20121560089 Lots affected OUS: BP20121510086, BP20121500080

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2074-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2074-2012Class IITerminatedVoluntary: Firm initiated