Siemens syngo.plaza. Radiological image processing system
Siemens Medical Solutions USA, Inc began a recall of Siemens syngo.plaza. Radiological image processing system on . FDA classified it as Class II and gives the reason as "Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2078-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens syngo.plaza. Radiological image processing system
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error message is displayed to user.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a Customer Safety Advisory Notice letter dated February 24, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to ensure that the prefix defined in the viewer settings for "Content Creator's Initials" is not empty. Then all images will be display correctly. This has to be checked for every user. Further questions please call 610-219-6300.
- Quantity in commerce
- 2
- Distribution
- (USA) nationwide including the states of CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
- Product codes and lots
- Model number 10592457
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2078-2012 | Class II | Terminated | Voluntary: Firm initiated |