Z-2210-2012Class II

Maquet Getinge Group Pediatric Arterial Cannula Maquet Cardiopulmonary AG Kehler Strabe 31 76437 Rastatt...

Maquet Cardiovascular Us Sales, LLC / initiated 2012-05-29 / Terminated / root cause: design

Maquet Inc. began a recall of Maquet Getinge Group Pediatric Arterial Cannula Maquet Cardiopulmonary AG Kehler Strabe 31 76437 Rastatt Germany Phone +49 7222 932-0 Fax: +49 7222 932-1888... on . FDA classified it as Class II and gives the reason as "Certain pediatric arterial cannulas may have limited flexibility due to an increased length of the "hardened" cannula tip". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2210-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Getinge Group Pediatric Arterial Cannula Maquet Cardiopulmonary AG Kehler Strabe 31 76437 Rastatt Germany Phone +49 7222 932-0 Fax: +49 7222 932-1888 [email protected] www.maquet.com Used to introduce blood and sterile solution to the circulation system during extracorporeal circulation procedures
Product code
NCP
Reason for recall
Certain pediatric arterial cannulas may have limited flexibility due to an increased length of the "hardened" cannula tip.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet issued Urgent Device 2, Recall letters/Urgent Product Recall Forms dated May 29, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their current stock and identify any affected product and remove from their inventory. Customers were instructed to contact the customer service department at 800-777-4222 (option #2, followed by option #2). Customers were also instructed to completed the enclsoed Recall Form to acknowledge receipt of the notification and to document the current status of the devices shipped to their facility. For questions regarding this recall call 973-709-7660.
Quantity in commerce
20 devices US; 122 devices foreign
Distribution
Worldwide Distribution - USA including MI, MN and Internationally
Product codes and lots
510k023251  US Catalog Number 161412 Batch Number 92048923   Foreign Catalog number 161412 Batch number 92048923 Catalog number 161412 Batch number 92051578

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2210-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2210-2012Class IITerminatedVoluntary: Firm initiated