Z-1950-2012Class II
Drager PT 4000 Phototherapy System
Draeger Medical Systems, Inc. / initiated 2012-07-02 / Terminated
Draeger Medical Systems, Inc. began a recall of Drager PT 4000 Phototherapy System on . FDA classified it as Class II and gives the reason as "A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1950-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger Medical Systems, Inc.
- Product
- Drager PT 4000 Phototherapy System
- Product code
- LBI Unit, Neonatal Phototherapy
- Reason for recall
- A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b
- FDA root cause
- Other
- Action
- Drager Medical Systems sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 2012 to all affected customers. The letter identifies the product, problem, and actions to bee taken by the customers. In addition, a Customer Reply Card was included for customers to complete and return. Contact the firm at 800-543-5047 for questions regarding this recall.
- Quantity in commerce
- 65
- Distribution
- Nationwide Distribution-USA (nationwide) and Puerto Rico.
- Product codes and lots
- Catalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1950-2012 | Class II | Terminated | Voluntary: Firm initiated |