Z-2397-2012Class II

BD FACS 7-color Setup beads BD Biosciences, San Jose, CA 95131. BD FACS 7 color Setup Beads are for in vitro...

Becton, Dickinson and Company, BD Biosciences / initiated 2012-06-29 / Terminated / root cause: manufacturing process

BD Biosciences, Systems & Reagents began a recall of BD FACS 7-color Setup beads BD Biosciences, San Jose, CA 95131. BD FACS 7 color Setup Beads are for in vitro diagnostic use on BD FACSCanto or BD FACSCanto II... on . FDA classified it as Class II and gives the reason as "Incorrect version of the CD was included with BD FACS 7 Color Setup Beads". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2397-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD FACS 7-color Setup beads BD Biosciences, San Jose, CA 95131. BD FACS 7 color Setup Beads are for in vitro diagnostic use on BD FACSCanto or BD FACSCanto II systems. The beads are run with BD FACSCanto clinical software. This system provides application specific setup, including automated compensation and cytometer quality control (QC) with minimal user intervention.
Product code
GKZ Counter, Differential Cell
Reason for recall
Incorrect version of the CD was included with BD FACS 7 Color Setup Beads.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
BD Biosciences sent a Important Product Recall Information letter beginning July 3, 2012, to all affected consignees via Fed Express. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed that the CD enclosed in their recent purchase of BD FACS 7-Color Setup Beads, lot number 33412, contain CD version 2.0 instead of the newer version 3.0. The letter goes on to explain the the potential risk and include a table of listed range values to determine if a potential false pass or false fail for the listed detectors has occurred. Customers are asked to check the CD with the kit to ensure that it is version 3.0 (the kit lot number ID is on the CD label). If it is not the correct version, discard version 2.0 and do not use it with this bead kit. Call BAD customers service for a replacement. If the 2.0 kit was already uses, letters request that clinicians inspect the FITC and APC-CY7 stains in the plots of the laboratory reports for the clinical applications in BD FACSCanto clinical software for compromised resolution between negative and dim populations. The verification form/effectiveness check should be completed and returned as requested. Questions or concerns may be directed to BD Customer Support at 1-855-236-2772.
Quantity in commerce
96 kits
Distribution
Worldwide Distribution -- USA (nationwide) including the countries of Australia, Belgium, Canada, Chile, Korea, Mexico, Singapore and Taiwan.
Product codes and lots
Catalog number: 335775; Lot number 33412.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2397-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2397-2012Class IITerminatedVoluntary: Firm initiated