Z-2187-2012Class II

The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable...

Zimmer GmbH / initiated 2012-07-13 / Terminated / root cause: labeling and UDI

Zimmer, Inc. began a recall of The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indicated for temporary internal fixation and... on . FDA classified it as Class II and gives the reason as "Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2187-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
The NCB Polyaxial Locking Plate System NCB screws 26mm: 02.03155.026, non-sterile Product is implantable screws indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones. These screws are intended for bicortical anchorage.
Product code
HRS Plate, Fixation, Bone
Reason for recall
Zimmer investigation following a complaint found one lot of NCB Polyaxial Locking Plate System NCB screws, 26mm, distributed in the United States contained incorrect 32 mm screws.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
The firm, Zimmer, sent an "URGENT MEDICAL DEVICE RECALL" letter dated July 13, 2012, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to locate the implicated lot in their inventory and user accounts; return all packaged units of this lot; and provide Zimmer with the name, complete mailing address, and phone number of each user facility and surgeon that have used the product via E-mail to [email protected]. Zimmer will send a follow-up Urgent Recall notification to these sub-accounts with instructions to verify the screw length of any screws already in the implant rack inventory prior to use in surgery. All implicated product is to be returned. The customers were also instructed to complete and return the Inventory Return Certification form to Zimmer via fax at +1-574-372-4265. If you have any questions about this correction and/or product removal, contact Zimmer at +1-877-946-2761.
Quantity in commerce
24 to US Distributor with 7 further distributed
Distribution
Nationwide distribution: USA including states of: AZ, IL, IN, ME and NJ.
Product codes and lots
Part number 02.03155.026, lot 2649941

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2187-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2187-2012Class IITerminatedVoluntary: Firm initiated