Z-2135-2012Class II

RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk)

Stryker Corp. / initiated 2012-07-17 / Terminated / root cause: labeling and UDI

Stryker Instruments Div. of Stryker Corporation began a recall of RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF... on . FDA classified it as Class II and gives the reason as "Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2135-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
RF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The Stryker RF Cannulae, in combination with the Stryker RF Generator/Multi- Gen and electrodes, are intended for coagulation of soft tissue in orthopedic, arthroscopic, spinal, and neurosurgical applications.
Product code
GXI Probe, Radiofrequency Lesion
Reason for recall
Stryker has received 6 complaints that the 5mm needle was placed in 10 mm package and 10 mm needle was placed in the 5 mm package. If the clinician does not recognize that the active tip on the cannula is 10 mm instead of 5mm and the cannula is used in a cervical procedure there is a potential for a thermal skin lesion due to a larger lesion creation. If the 10 mm active tip cannula, as opposed
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Stryker sent an Urgent MEDICAL DEVICE Recall Notifications dated July 16, 2012 to all consignees. The letter identified the affected products, problem and actions to be taken. Customers/distributors were instructed to check their inventory and quarantine any affected product found, complete and fax the business reply form to Stryker Instruments Regulatory Department, 866-521-2762 and return all affected product using the pre-paid shipper provided. Upon receipt of the recalled product, a replacement will be issued. For questions regarding this recall please contact Stryker Instruments (Recall Coordinator) Monday-Friday 8am-5pm (EST) at 269-389-4354.
Quantity in commerce
226
Distribution
Worldwide Distribution - US Nationwide and the countries of England, France, Netherlands, and Switzerland.
Product codes and lots
REF 0406-630-125, lot 1000037535

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2135-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2134-2012Class IITerminatedVoluntary: Firm initiated
Z-2135-2012Class IITerminatedVoluntary: Firm initiated