Z-2121-2012Class II

Perkin Elmer Genetic Screening Processor (GSP)

Revvity Health Sciences, Inc. / initiated 2012-06-27 / Terminated / root cause: design

Perkinelmer began a recall of Perkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a fully automated, high throughput batch analyzer... on . FDA classified it as Class II and gives the reason as "Contamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2121-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Revvity Health Sciences, Inc.
Product
Perkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a fully automated, high throughput batch analyzer for time- resolved and prompt fluorescence analysis of samples in microtitration plates. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids. The reagent kits running on the GSP instrument are used to test blood specimens dried on filter paper as an aid in screening newborns for congenital disorders. Currently assays for the determination of T4, TSH, 17OHP, IRT and GALT are cleared and available for use with the GSP instrument.
Product code
JLW Radioimmunoassay, Thyroid-Stimulating Hormone
Reason for recall
Contamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT
Root cause tag
design
FDA root cause
Component design/selection
Action
PerkinElmer sent an Urgent: Medical Device Recall letter dated June 27, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter states that replacement units of the recalled product will be installed as soon as they are available. Customers were instructed to complete and return the response form attached. For further information contact your local PerkinElmer representative or GS. [email protected]
Quantity in commerce
78 units
Distribution
Worldwide Distribution - US (nationwide) and the countries of Australia, Belgium, Brazil, Egypt, France, Germany, Netherlands, Norway, Saudi Arabia,Switzerland,
Product codes and lots
Serial Numbers: 20210001 through 20210078

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2121-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2121-2012Class IITerminatedVoluntary: Firm initiated