Z-2317-2012Class II
BD GeneOhm MRSA ACP Assay, Catalog #441639, box 200 tests labeled in part***GeneOhm Sciences Canada, Inc...
Becton Dickinson & Co / initiated 2012-05-22 / Terminated / root cause: manufacturing process
Becton Dickinson & Co. began a recall of BD GeneOhm MRSA ACP Assay, Catalog #441639, box 200 tests labeled in part***GeneOhm Sciences Canada, Inc. 2555 Boul, du Parc-Technologique, Quebec, QC, Canada... on . FDA classified it as Class II and gives the reason as "Leakage in Cepheid SmartCycler reaction tubes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2317-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD GeneOhm MRSA ACP Assay, Catalog #441639, box 200 tests labeled in part***GeneOhm Sciences Canada, Inc. 2555 Boul, du Parc-Technologique, Quebec, QC, Canada G1P 4S5*** BD GeneOhm MRSA ACP Assay is a qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization. Test is intended to aid in the prevention and control of MRSA infections in healthcare settings. It is not intended to diagnose MRSA infections nor to guide or monitor treatment for MRSA infections. .
- Product code
- NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen
- Reason for recall
- Leakage in Cepheid SmartCycler reaction tubes.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- BD Diagnostic Systems issued an Urgent Product Recall letter via UPS overnight delivery on May 22, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following actions: Discontinue use and discard any affected product. BD would send replacement tubes for the discarded material. Complete the attached form whether or not they have any inventory remaining so that the firm may acknowledge their receipt of the notification. Fax the form to: Attention: Regulatory Compliance FAX: 410-316-4258 Email: [email protected] If further assistance is needed, customers should contact BD Technical Services Department at 1-800-838-8663. For questions regarding this recall call 410-316-4054.
- Quantity in commerce
- 68 kits
- Distribution
- Nationwide Distribution including CA, DE, FL, GA, IL, IN, KS, KY, ME, MT, NC, NJ, NM, NY, OH, OK, PA, SC, TX, VA, WA, WI, and WV.
- Product codes and lots
- Lot/ Exp. 03T11345ZU 2012-07-09
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2316-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2317-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2318-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2319-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2320-2012 | Class II | Terminated | Voluntary: Firm initiated |