Z-2167-2012Class II

Smith & Nephew 5.5 mm TWINFIX Ultra Ti Anchor with two #2 sutures (COBRAID-blue, COBRAIDblack), sterile Part...

Smith & Nephew, Inc. / initiated 2012-07-03 / Terminated / root cause: design

Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew 5.5 mm TWINFIX Ultra Ti Anchor with two #2 sutures (COBRAID-blue, COBRAIDblack), sterile Part Number: 72202896 Product Usage: intended for use... on . FDA classified it as Class II and gives the reason as "Distal part of the anchor may break on insertion into bone during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2167-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew 5.5 mm TWINFIX Ultra Ti Anchor with two #2 sutures (COBRAID-blue, COBRAIDblack), sterile Part Number: 72202896 Product Usage: intended for use for the reattachment of soft tissue to bone
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Distal part of the anchor may break on insertion into bone during surgery
Root cause tag
design
FDA root cause
Device Design
Action
Smith & Nephew sent an Urgent Product Recall 1st Notification letters dated June 2012 via Federal Express to all affected customers. The letter identified the affected product, potential risk and actions to be taken. Customers were instructed to inspect inventory and quarantine them immediately, complete and return the attached form indicating the Lot# and quantities that need to be returned and contact Smith & Nephew Returns Group at 800-343-5717 (option 3) or email to [email protected]. for instructions for returning affected products and receiving a credit. For questions call 1-508-261-3731.
Quantity in commerce
2,204 units
Distribution
Worldwide Distribution - US Nationwide and the countries of: Argentina, AUSTRALIA, AUSTRIA, Belgium, CANADA,CHILE COLOMBIA, Denmark, DUBAI, FINLAND, FRANCE GDC BAAR Distribution,GERMANY, GREECE, INDIA ITALY, KOREA, MALAYSIA, MEXICO, NETHERLAND, NORWAY PORTUGAL, PUERTO RICO, SHANGHAI, SOUTH AFRICA SPAIN, SWEDEN, SWITZERLAND, and UNITED KINGDOM.
Product codes and lots
Lot Numbers: 50388182, 50389253, 50391626, 50392272, 50392407, 50395081, 50395680, 50396197, 50400556, 50401558, 50402228, 50403940, 50406036, 50407582, 50409406, 50413407, 50414970, 50412647, 50415864, 50416801, 50418088, 50420654, 50421030, 50421495

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2167-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2165-2012Class IITerminatedVoluntary: Firm initiated
Z-2166-2012Class IITerminatedVoluntary: Firm initiated
Z-2167-2012Class IITerminatedVoluntary: Firm initiated
Z-2168-2012Class IITerminatedVoluntary: Firm initiated