Z-1993-2012Class II

GE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to take...

GE Healthcare / initiated 2012-04-04 / Terminated / root cause: software

GE Healthcare, LLC began a recall of GE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to take exposures utilizing film or computed radiography (... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a software issue on the interface of the mobile radiographic product, Optima XR200amx, ond Optima XR220amx related to Visual indication as mandated by the US Code of Federal Regulations...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1993-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to take exposures utilizing film or computed radiography ( CR )
Product code
IZL System, X-Ray, Mobile
Reason for recall
GE Healthcare has become aware of a software issue on the interface of the mobile radiographic product, Optima XR200amx, ond Optima XR220amx related to Visual indication as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-1-3.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to the user or through a GE Healthcare field engineer site visit. A GE Healthcare Service Representative will update the firmware on the system to address the issue. Your proposed corrective action plan (CAP) (Field Modification Instructions 10865) describes the rework plan that GEHC Engineers will make to the imaging system to return it to compliance with the applicable performance standard. It appears to adequately address the problem and is hereby approved.
Quantity in commerce
254 units have been shipped to/installed in the U.S.
Distribution
Worldwide Distribution (USA) nationwide
Product codes and lots
Model Numbers Optima XR200 15kw Optima XR 200 30kw Optima XR 220 15kw Optima XR 220 30kw

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1993-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1993-2012Class IITerminatedVoluntary: Firm initiated