I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for...
Abbott Point Of Care Inc. began a recall of I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation... on . FDA classified it as Class II and gives the reason as "The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2157-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.
- Product code
- GJS Test, Time, Prothrombin
- Reason for recall
- The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Abbott Point of Care Inc, sent a URGENT RECALL NOTICE dated March 2012 , to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were provided in the letter with instructions in the letter on how to handle the affected product. Please complete and return the enclosed Business Reply card to acknowledge receipt of this communication and installation of the interium software CLEW and also 1) Update all handhelds 2) download appropriate assignment sheets. For further questions please call (609) 454-9000.
- Quantity in commerce
- 2,950,296 cartridges US; 156,408 cartridges OUS
- Distribution
- Worldwide Distribution -- (USA) nationwide
- Product codes and lots
- List number 04J50-01 04J5O-02 03P89-24 Lots begininng with R11 or R12
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2157-2012 | Class II | Terminated | Voluntary: Firm initiated |