Z-2157-2012Class II

I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for...

Abbott Medical / initiated 2012-03-01 / Terminated / root cause: manufacturing process

Abbott Point Of Care Inc. began a recall of I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation... on . FDA classified it as Class II and gives the reason as "The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2157-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.
Product code
GJS Test, Time, Prothrombin
Reason for recall
The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Abbott Point of Care Inc, sent a URGENT RECALL NOTICE dated March 2012 , to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were provided in the letter with instructions in the letter on how to handle the affected product. Please complete and return the enclosed Business Reply card to acknowledge receipt of this communication and installation of the interium software CLEW and also 1) Update all handhelds 2) download appropriate assignment sheets. For further questions please call (609) 454-9000.
Quantity in commerce
2,950,296 cartridges US; 156,408 cartridges OUS
Distribution
Worldwide Distribution -- (USA) nationwide
Product codes and lots
List number 04J50-01 04J5O-02 03P89-24  Lots begininng with R11 or R12

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2157-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2157-2012Class IITerminatedVoluntary: Firm initiated