94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing
ICU Medical, Inc. began a recall of 94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing on . FDA classified it as Class II and gives the reason as "The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2130-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- 94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing
- Product code
- FMG Stopcock, I.V. Set
- Reason for recall
- The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- ICU Medical sent an Urgent Medical Device Recall letter dated April 2012 to all affected customers. The letter identified the recalled product, problem and actions to be taken. Customers were instructed to inspect inventory, destroy affected products according to hospital protocol, complete and return verification form and contact ICU Customer Service to order replacement products. The letter states that ICU Medical will credit customer for any product destroyed once the completed verification form is received. This recall should be carried out to the user level and passed on to all those who need to be aware within your organization or any other organization the device may have been transferred to. Should you have any questions or require assistance relating to this recall please contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time: 866-829-9025 and select option 8, or e-mail the following address: [email protected]
- Distribution
- Nationwide distribution
- Product codes and lots
- Part Number B30059 510(k) K964435 Lot number 2418897, 2423996, 2425137, 2430044, 2431861, 2436569, 2443559, 2447238, 2448099, 2452377, 2457453, 2459025, 2464569, 2467687.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2128-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2129-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2130-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2131-2012 | Class II | Terminated | Voluntary: Firm initiated |