Z-2159-2012Class III

Cobas KRAS Mutation Test for In Vitro Diagnostic Use

Roche Molecular Systems Inc / initiated 2012-02-24 / Terminated / root cause: manufacturing process

Roche Molecular Systems, Inc. began a recall of Cobas KRAS Mutation Test for In Vitro Diagnostic Use Product Usage: Usage: The Cobas KRAS Mutation Test, for use with the Cobas 4800 System, is a real-time PCR... on . FDA classified it as Class III and gives the reason as "During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was identified that the kit was just within specifications for the codon 61 peak height ratio, and that the kit...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2159-2012
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Molecular Systems Inc
Product
Cobas KRAS Mutation Test for In Vitro Diagnostic Use Product Usage: Usage: The Cobas KRAS Mutation Test, for use with the Cobas 4800 System, is a real-time PCR test intended for the identification of mutations in codons 12, 13 and 61 of the KRAS gene in DNA derived from formalin-fixed paraffin-embedded human colorectal cancer (CRC).
Product code
OOI Real Time Nucleic Acid Amplification System
Reason for recall
During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was identified that the kit was just within specifications for the codon 61 peak height ratio, and that the kit has trended towards the upper specification limit at the 6 month time point but did passed the 6 month stability testing. The kit is labeled for 9 months shelf life stability. The d
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Roche issued a Safety Board Notice (Prooduct Support Notice 2012-01) and Product Bulletin 2012/03 dated February 24, 2012 to all affected customers. The notice idenfied the affected product, description of situation and actions to be taken. Customers were instructed to discontinue use of lot P06778 by March 4, 2012 and to quarantine and discard any remaining kits from lot P06778. The notice stated that a new lot will be made available as a replacement.
Quantity in commerce
264 kits to 13 EU countries; 2 kits to Canada and 33 kits to 11 countries in the Rest of the World
Distribution
Worldwide Distribution - including the countries of Austria, Belgium, Czech Republic, Denmark, Germany, Italy, Luxemburg, Netherlands, Poland, Slovakia, Slovenia, Spain, United Kingdom, Canada, Chile, Greenland, Hong Kong, Israel, Malaysia, New Zealand, Philippines, S. Africa, Switzerland, Taiwan and Turkey.
Product codes and lots
05852170190; Lot P06778

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2159-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2159-2012Class IIITerminatedVoluntary: Firm initiated