Z-2156-2012Class II

Philips MultiDiagnost Eleva with Flat Detector (MD Eleva FD)Model: 708034

Philips North America LLC / initiated 2012-07-11 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Philips MultiDiagnost Eleva with Flat Detector (MD Eleva FD)Model: 708034 Product Usage: Philips X-ray System on . FDA classified it as Class II and gives the reason as "When doing a 3 Dimensional Rotational Angiography (3DRA) the 3D reconstruction does not always start, even when all images have been acquired". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2156-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips MultiDiagnost Eleva with Flat Detector (MD Eleva FD)Model: 708034 Product Usage: Philips X-ray System
Product code
KPR System, X-Ray, Stationary
Reason for recall
When doing a 3 Dimensional Rotational Angiography (3DRA) the 3D reconstruction does not always start, even when all images have been acquired.
Root cause tag
software
FDA root cause
Software design
Action
Philips issued an Urgent Field Safety Notice dated July 2, 2012 to all affected customers. The letter identified the affected product, problem description, actions to be taken by the customer/user in order to prevent risks for patients or users and actions planned by Philips to correct the problem. For question or support concerning this issue contact your local Philips representative.
Quantity in commerce
41 units
Distribution
Worldwide Distribution - US (nationwide) including the states of CA, CT, MI, MN, MN, NC, NY, NY, VA and the countries of Belgium, Columbia, Czech Republic and Denmark. France Germany Italy Latvia Luxembourg Netherlands Russian Fed. Spain Sweden Switzerland Thailand United Kingdom
Product codes and lots
US Serial Numbers: 96 104 111 120 170 222 223 272 278 WW: 41 70 83 96 104 111 120 132 154 155 156 157 167 170 171 176 179 190 194 195 200 208 216 222 223 224 237 244 245 247 248 249 251 253 259 261 262 264 265 272 278

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2156-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2156-2012Class IITerminatedVoluntary: Firm initiated