Z-2150-2012Class II

Finapres Medical Systems Portapres with NiMh battery pack

Finapres Medical Systems B.V. / initiated 2012-07-06 / Terminated / root cause: manufacturing process

FMS, Finapres Medical Systems BV began a recall of Finapres Medical Systems Portapres with NiMh battery pack Product Usage: Continuously measure blood pressure, The Portapres can be worn on the body with a... on . FDA classified it as Class II and gives the reason as "Battery may become hot and leak". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2150-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Finapres Medical Systems B.V.
Product
Finapres Medical Systems Portapres with NiMh battery pack Product Usage: Continuously measure blood pressure, The Portapres can be worn on the body with a protective waist belt. The device can be used without mains; in that case a battery pack is used.
Product code
DXN System, Measurement, Blood-Pressure, Non-Invasive
Reason for recall
Battery may become hot and leak
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Finapres Medical Systems BV sent a Field Safety Notice letter dated July 6, 2012 to notified US accounts of the recall. The letter identified the affected product, description of the problem and actions to be taken. The letter states that the recalling firm recommends not using the affected battery packs. A replacement battery was included in shipment. Customers were requested to return recalled battery. For questions call +31 (0)20 609 0974 or email [email protected].
Quantity in commerce
3 units US
Distribution
US Nationwide Distribution - including the states of: OH and NY
Product codes and lots
Serial Numbers: 12 02 006, 12 02 008, 12 02 013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2150-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2150-2012Class IITerminatedVoluntary: Firm initiated