Z-2140-2012Class I

AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders

Baxter Healthcare Corporation / initiated 2012-07-13 / Terminated / root cause: design

Baxter Healthcare Corp. began a recall of AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical... on . FDA classified it as Class I and gives the reason as "After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2140-2012
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.
Product code
NEP System/Device, Pharmacy Compounding
Reason for recall
After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
Root cause tag
design
FDA root cause
Device Design
Action
Baxter Healthcare Corp. sent an Urgent Product Recall letters dated July 13, 2012, via first class mail to all affected customers, informing personnel that Baxter was executing a voluntary controlled product withdrawal of the Automix system from the U.S. market by June 1, 2013. Until their units can be replaced, customers were advised to exercise care when entering data, and verify the accuracy of the keypad response; to inspect the umbilical cable of the Automix for any visible cuts, nicks deformation, or physical damage to the cable, discontinuing use of the device if such damage is found; and to return the device to Baxter if the device stops compounding without an alarm and without a visible green complete light, discarding the solution bag. The customers were requested to convey this information to all users of the Automix compounders, and a wall chart was provided to hang over the device to remind users of the mitigations to take. The customers were also requested to complete the attached customer reply form listing the model and serial number of each of the Automix compounders at the facility and confirming their receipt of the letter, and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to their local Baxter representative, and technical questions were directed to Baxter's Global Technical Services at 1-800-626-2667.
Quantity in commerce
840 units
Distribution
Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.
Product codes and lots
Product code 2M8075, all serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2140-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2139-2012Class ITerminatedVoluntary: Firm initiated
Z-2140-2012Class ITerminatedVoluntary: Firm initiated
Z-2141-2012Class ITerminatedVoluntary: Firm initiated
Z-2142-2012Class ITerminatedVoluntary: Firm initiated