Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2146-2012Class II

djo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell...

Encore Medical, LP / initiated 2012-07-16 / Terminated / root cause: packaging

Encore Medical, Lp began a recall of djo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell or fracture repair on . FDA classified it as Class II and gives the reason as "One lot of the 010-55-030, 6.5 mm Cancellous Bone Screw was recalled because a product complaint identified a mislabeling of the device. Although the package label indicates Size 30mm, the screw dimension is actually...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2146-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
djo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell or fracture repair
Product code
HWC Screw, Fixation, Bone
Reason for recall
One lot of the 010-55-030, 6.5 mm Cancellous Bone Screw was recalled because a product complaint identified a mislabeling of the device. Although the package label indicates Size 30mm, the screw dimension is actually 25mm.
Root cause tag
packaging
FDA root cause
Packaging
Action
DJO Global sent an Urgent Field Safety Notice dated July 16, 2012, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to forward the notice to all thoe who need to be aware within their organization or to any other organization where the potentially affected product was transferred, contact customers to place a replacement order and receive an RMA number, and return all affected devices using the RMA number. For questions customers should call 512-834-6255. For questions regarding this recall call 512-832-9500.
Quantity in commerce
18 devices
Distribution
Nationwide Distribution including AR, CA, FL, ID, OK, PA, TX, and UT.
Product codes and lots
Lot 007A1037, Ref 010-55-25.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2146-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2146-2012Class IITerminatedVoluntary: Firm initiated