Z-2153-2012Class II

Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA...

Varian Medical Systems, Inc. / initiated 2012-07-02 / Terminated / root cause: software

Varian Medical Systems, Inc. Oncology Systems began a recall of Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is... on . FDA classified it as Class II and gives the reason as "A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2153-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.
Root cause tag
software
FDA root cause
Software design
Action
Varian Medical Systems sent a " Urgent Medical Device Correction Urgent Field Safety Notice" dated July 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed until a correction is deployed all treatment plans using the BrainLab m3 mMLC must be imported using the Import Wizard only. Varian Medical Systems is developing a technical correction for this issue. You will be contacted by a Customer Service Representative when this correction is available to schedule its installation on your system. For further questions please call 1.888.827.4265.
Quantity in commerce
2 Units.
Distribution
Nationwide Distribution including the states of New York and Ohio.
Product codes and lots
Serial Numbers/CODES: HIT0535, HIT2953

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2153-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2153-2012Class IITerminatedVoluntary: Firm initiated