Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA...
Varian Medical Systems, Inc. Oncology Systems began a recall of Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is... on . FDA classified it as Class II and gives the reason as "A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2153-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Varian Medical Systems sent a " Urgent Medical Device Correction Urgent Field Safety Notice" dated July 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed until a correction is deployed all treatment plans using the BrainLab m3 mMLC must be imported using the Import Wizard only. Varian Medical Systems is developing a technical correction for this issue. You will be contacted by a Customer Service Representative when this correction is available to schedule its installation on your system. For further questions please call 1.888.827.4265.
- Quantity in commerce
- 2 Units.
- Distribution
- Nationwide Distribution including the states of New York and Ohio.
- Product codes and lots
- Serial Numbers/CODES: HIT0535, HIT2953
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2153-2012 | Class II | Terminated | Voluntary: Firm initiated |