Z-2208-2012Class II

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032...

Philips North America LLC / initiated 2012-07-11 / Terminated / root cause: manufacturing process

Philips Healthcare Inc. began a recall of Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for... on . FDA classified it as Class II and gives the reason as "Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2208-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.
Product code
IXR Table, Radiographic, Tilting
Reason for recall
Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Philips issued an Urgent Field Safety Notice dated July 9, 2012, to notify users that the inclination swtich may be mounted incorrectly. A Field Safety Engineer will be dispatched to verify if a switch was installed properly, and in the event it was not installed properly they will re-install it properly. The Field Service Engineer will add a label as described in the spare part replacement manual, which is part of the FCO kit.. Customers were instructed review the information contained in the Field Safety Notice and to retain a copy of the notice with the equipment Instruction for Use. For further questions customers were instructed to call their Phillips representataive Technical Support Line at 1-800-722-9377. For questions regarding this recall call 978-687-1501
Quantity in commerce
12 units
Distribution
Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of Canada, HongKong, New Zealand.
Product codes and lots
Serial Numbers: SN07000141 SN07000109 SN5000162 SN5000421 SN07000667 200148 200115 200081 300114

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2208-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2208-2012Class IITerminatedVoluntary: Firm initiated