Z-2244-2012Class II
Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120...
Medtronic Navigation, Inc / initiated 2012-07-11 / Terminated
Medtronic Navigation, Inc. began a recall of Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R on . FDA classified it as Class II and gives the reason as "A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2244-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.
- FDA root cause
- Component change control
- Action
- Medtronic Navigation issued an "Urgent Field Safety Notice" to consignees dated 7/11/12. The problem was described and recommended actions were provided. For each affected O-arm Imaging System, a Medtronic service representative will visit the site and replace the 500VDC relay on the system and test the O-arm Imaging System.
- Quantity in commerce
- 17 units
- Distribution
- Worldwide Distribution - USA, including the states of AL, CO, FL, IA, MI, IA, NC, NE, NY and the countries of South Africa, Italy, and Spain.
- Product codes and lots
- Serial Numbers: 109R, 118R, 124, 136, 159 ,170, 218, 222, 223, 236, 239 , 299, 309R, 402, 410, 412, 414.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2244-2012 | Class II | Terminated | Voluntary: Firm initiated |